Overview
Spot-check or continuously monitor three critical parameters with a single monitor. Designed to travel everywhere you do, the VitalCareTM 506N3 Series with integrated carrying handle attaches quickly and easily to roll stands. Available in multiple configurations, the VitalCare provides maximum versatility and cost-effectiveness.
Smart battery management allows up to 100 non-invasive blood pressure measurements on a single battery charge. With a compact design that is easily set up for spot-checks or continuous monitoring, the monitor adapts readily to hospital, clinic and transport requirements.
Features
- ComfortCuff® NIBP measures on-inflation for faster, more comfortable readings.
- Choice of CSI’s DOX® or Nellcor OxiMax® pulse oximetry.
- Audible and visual alarms are adjustable for use in continuous monitoring.
- Large responsive keys for fast spot-checks with easy to clean membrane keypad.
- Accomodates all patient types from adult to neonate.
- Multi-patient trend memory.
- Rechargeable battery with direct AC power.
- Optional FILAC or Alaris TurboTemp® predictive temperature for quick, accurate measurements.
- Optional integrated printer for on-demand, automatic interval, trends or multipatient reports.
Options
Part Number | Description |
506N3 | NIBP, Heart Rate |
506NP3 | NIBP, Heart Rate, Printer |
506NT3 | NIBP, Heart Rate, FILAC Temperature |
506NTP3 | NIBP, Heart Rate, FILAC Temperature, Printer |
506DN3 | NIBP, Heart Rate, DOX SpO2 |
506DNP3 | NIBP, Heart Rate, DOX SpO2, Printer |
506DNT3 | NIBP, Heart Rate, DOX SpO2, FILAC Temperature |
506DNTP3 | NIBP, Heart Rate, DOX SpO2, FILAC Temperature, Printer |
506LN3 | NIBP, Heart Rate, OxiMax SpO2 |
506LNP3 | NIBP, Heart Rate, OxiMax SpO2, Printer |
506LNT3 | NIBP, Heart Rate, OxiMax SpO2, FILAC Temperature |
506LNTP3 | NIBP, Heart Rate, OxiMax SpO2, FILAC Temperature, Printer |
506NV3 | NIBP, Heart Rate, TurboTemp Temperature |
506NVP3 | NIBP, Heart Rate, TurboTemp Temperature, Printer |
506DNV3 | NIBP, Heart Rate, DOX SpO2, TurboTemp Temperature |
506DNVP3 | NIBP, Heart Rate, DOX SpO2, TurboTemp Temperature, Printer |
506LNV3 | NIBP, Heart Rate, OxiMax SpO2, TurboTemp Temperature |
506LNVP3 | NIBP, Heart Rate, OxiMax SpO2, TurboTemp Temperature, Printer |
Each product comes with (1) NIBP hose, (1) child/small adult NIBP cuff, (1) adult NIBP cuff, (1) large adult NIBP cuff, (1) rechargeable battery, (1) power cord, and (1) user manual on CD-Rom.
Additional accessories based on configurations include: (1) SpO2 sensor, (1) oral temperature well and probe, (1) box of 20 temperature probe covers, (1) roll of printer paper.
Specifications
NIBP (NON-INVASIVE BLOOD PRESSURE)
Technique: | Oscillometric measure upon inflation |
Measurement Cycle: | < 40 sec. typical |
STAT Mode: | 5 min. of continuous measurements |
Automatic Measurement: | 1, 2, 3, 5, 10, 15, 30, 45, 60 min.; 2, 4 hrs; STAT |
Inflation Pressure Range: | Adult: 30 to 300 mmHg; Pediatric: 30 to 150 mmHg; Neonate: 20 to 150 mmHg |
Resolution: | 1 mmHg |
ransducer Accuracy: | ±2 mmHg or ±2% of reading, whichever is greater over full range |
Clinical Accuracy: | Complies with AMI SP-10:2002 |
PULSE OXIMETRY
Range: | 1% to 99% |
Resolution: | 1% |
Accuracy (DOX only): | 70% to 99% range: ±2%; 50% to 69% range: ±3%; < 49%: unspecified |
Accuracy (OxiMax only): | 70% to 99% range: ±2% < 69%: unspecified |
Method: | Dual wavelength LED |
Audio: | Tone with each detected pulse; pitch varies with saturation |
HEART RATE (PULSE)
Range: | 20 to 300 bpm |
Accuracy (DOX Only): | ±1 bpm or ±1% of reading, whichever is greater; unspecified for <30 or >240 bpm |
Accuracy (OxiMax Only): | ±2 bpm or ±1% of reading, whichever is greater; unspecified for <30 or >240 bpm |
Update Time: | 1 sec. |
Resolution: | 1 bpm |
Pulse Volume: | 10 levels or Off |
TEMPERATURE (TURBOTEMP)
Type: | TurboTemp |
Method: | Predictive or continuous |
Channels: | 1 |
Predictive Mode Range: | 96 to 106 degrees F (35.6 to 41.1 degrees C) |
Continuous Mode Range: | 80 to 106 degrees F (26.7 to 41.1 degrees C) |
Accuracy: | ±0.2 degrees F (±0.1 degrees C; measured in calibrated water bath while in continuous mode |
Resolution: | ±0.1 degrees F or ±0.1 degrees C |
Response Time: | 10 seconds typical (oral) |
TEMPERATURE (FILAC)
Method: | Predictive of Continuous |
Channels: | 1 |
Range: | 86° to 109.4F (30° to 43°C |
Accuracy (Continuous): | ±0.2°F (±0.1°C) |
Accuracy: | (Predictive): ±0.2°F (±0.1°C) (Compliant with EN 12470-3 and ASTM E1112:98) |
Resolution: | ±0.1°F (±0.1°C) |
Predictive Response Time: | 10 sec. typical |
ALARMS
Indication: | Audible, Visual |
Volume: | 10 levels or OFF |
Silence: | Yes; 2 minutes or permanent |
TRENDS
Memory: | Last 24 hours |
Parameters: | NIBP, SpO2, pulse rate, temperature |
Report Types: | Single patient or multi-patient; selectable formats |
DISPLAY
Type: | LED |
Parameters: | SpO2, pulse rate, NIBP, temperature, pulse bar |
Status: | Alarm Silence; Sensor; Low Battery; External Power; Patient Size |
Text: | Alphanumeric LCD |
Languages: | English, French, German, Italian, Spanish, Portuguese, Dutch and Danish |
SYSTEM OUTPUTS
COM Port: | Serial, RS232 compatible |
PRINTER (OPTIONAL)
Type: | Thermal array |
Tabular Mode: | 1 to 60 minute intervals; 2 to 24 intervals; auto on BP; auto on alarm; off |
PHYSICAL PROPERTIES
Weight: | Base unit with battery: 6.95 lbs (3.1 kg) |
Size: | 8.7 in (H) x 7.1 in (W) x 6.5 in (D); 22 cm (H) x 18 cm (W) x 16.5 cm (D) |
POWER REQUIREMENTS
Voltage: | 100 to 240 VAC; 50/60 Hz |
Battery: | 1 sealed lead acid |
Battery Life: | 8.0 hours, typical |
Recharge Time: | 6.5 hours, typical |
ENVIRONMENTAL
Operating Temperature: | 32° to 113°F (0° to 45°C) |
Storage Temperature: | 23° to 122°F (-5° to 50°C) |
Operating Humidity: | 15% to 90%, noncondensing |
CLASSIFICATION
Electrical Protection: | Class I Equipment |
Degree of Protection: | Type CF, Defibrillator-proof |
Protection Against Ingress: | IPX1, Drip-proof Equipment |
MONITOR CONFIGURATIONS
Catalog Number: | Description: |
506N3 | NIBP only |
506NP3 | NIBP and Printer |
506NV3 | NIBP and Turbo Temp™ |
506NVP3 | NIBP, TurboTemp™, and Printer |
506DNP3 | NIBP and DOX SpO2 |
506DN3 | Standard with CO2, printer |
506DNV3 | NIBP, DOX SpO2, and TurboTemp™ |
506DNVP3 | NIBP, DOX SpO2, TurboTemp™ and Printer |
506LN3 | NIBP and Nellcor SpO2 |
506LNP3 | NIBP, Nellcor SpO2, and Printer |
506LNV3 | NIBP, Nellcor SpO2, and TurboTemp™ |
506LNVP3 | NIBP, Nellcor SpO2, TurboTemp™, and Printer |
506NT3 | NIBP and FILAC |
506NTP3 | NIBP, FILAC, and Printer |
506DNT3 | NIBP, DOX SpO2, and FILAC |
506DNTP3 | NIBP, DOX SpO2, FILAC, and Printer |
506LNT3 | NIBP, Nellcor SpO2, and FILAC |
506LNTP3 | NIBP, Nellcor SpO2, FILAC, and Printer |
Resources
Spec Sheets
OFFICE MANAGEMENT SOFTWARE COMPATIBILITY – CSI products are compatible with the following OMS products below. Contact software provider to confirm compatibility with specific configurations of Criticare monitors.
- Kodak WINOMS CS by Carestream Dental LLC
- TiME for OMS by Decision Base, Inc.
- U.S. HealthRecord powered by TeamLINKS, Inc.
- Oral Surgery-Exec® by DSN Practice Management Software
- OMSVision® by HenrySchein Practice Software
- Windent® by Carestream Dental LLC
- MD-REPORTS
- ProVation® Medical
- gMed
- Endosoft®
Contact Customer Service for details: 1-800-458-4615 (USA) 1-262-798-8282 (International)
Contact
U.S. CORPORATE HEADQUARTERS
333 Strawberry Field Rd Suite 11
Warwick, RI ,02886
T : 800 – 486 – 3849
Hours: 8:30 am to 5:30 pm Eastern Time